Acta Chir Orthop Traumatol Cech. 2006; 73(1):28-33 | DOI: 10.55095/achot2006/005

Měření otěru polyetylenových jamek TEP kyčelního kloubu univerzálním měřicím mikroskopem
Charakteristika měřeníPůvodní práce

J. GALLO1,*, V. HAVRÁNEK2, J. ZAPLETALOVÁ3, D. MANDÁT2
1 Ortopedická klinika LF UP a FN Olomouc
2 Společná laboratoř optiky UP a FZÚ AV ČR, Olomouc
3 Ústav biofyziky LF UP v Olomoci, pracoviště biometrie

PURPOSE OF THE STUDY:
The aim of the study was to present our own method of measuring acetabular polyethylene wear in total hip arthroplasty, including assessment of basic measurement characteristics. It is a well known fact that, in total hip replacement, polyethylene liners wear out with use. The rate of acetabular polyethylene wear can then be related to clinical or demographic data.

MATERIAL AND METHODS:
Our method, using a universal measuring microscope, is based on the ability of exact identification of central positions of the head of total hip replacement before implantation and after prosthesis removal. The difference between the original and final positions allows us to calculate both linear and volumetric wear of polyethylene. During one month, the same components were repeatedly measured by two independent investigators and a total of 10 ABG 1 acetabular components were checked. The results of measurements were evaluated by a series of statistical tests, including correlation and regression analyses and analysis of variance.

RESULTS:
High correlations were found among individual measurements made by each observer (r = 0.998; r = 0.973) as well as between the mean values obtained from the two observers (r = 0.996). The reliability of measurements was proved by a high correlation of the regression curve of each measurement with the "ideal" line. With two exceptions, the differences between paired measurements were not significant.

CONCLUSIONS:
Our method facilitates an in vitro measurement of polyethylene wear with considerable accuracy and high reliability.

Klíčová slova: total hip arthroplasty, wear measurement, linear wear, volumetric wear, polyethylene, universal measuring microscope, intra-observer variability, inter-observer variability

Zveřejněno: 1. únor 2006  Zobrazit citaci

ACS AIP APA ASA Harvard Chicago Chicago Notes IEEE ISO690 MLA NLM Turabian Vancouver
GALLO J, HAVRÁNEK V, ZAPLETALOVÁ J, MANDÁT D. Měření otěru polyetylenových jamek TEP kyčelního kloubu univerzálním měřicím mikroskopem
Charakteristika měření. Acta Chir Orthop Traumatol Cech. 2006;73(1):28-33. doi: 10.55095/achot2006/005. PubMed PMID: 16613745.
Stáhnout citaci

Reference

  1. BERZINS, A., SUMNER, D. R., GALANTE, J. O.: Dimensional characteristics of uncomplicated autopsy-retrieved acetabular polyethylene liners by ultrasound. J. Biomed. Mater. Res., 39: 120-129, 1998. Přejít k původnímu zdroji...
  2. ČECH, O., DŽUPA, V.: The European school of total hip arthroplasty and 35 years of total hip arthroplasty in the Czech Republic. Acta Chir. orthop. Traum. čech., 72: 57-76, 2005.
  3. DERBYSHIRE, B.: The estimation of acetabular cup wear volume from two-dimensional measurements: a comprehensive analysis. Proc. Inst. Mech. Eng. [H], 212: 281-291, 1998. Přejít k původnímu zdroji... Přejít na PubMed...
  4. DUMBLETON, J. H., MANLEY, M. T., EDIDIN, A. A.: A literature review of the association between wear rate and osteolysis in total hip arthroplasty. J. Arthroplasty, 17: 649-661, 2002. Přejít k původnímu zdroji... Přejít na PubMed...
  5. EBRAMZADEH, E., SANGIORGIO, S. N., LATTUADA, F., KANG, J. S., CHIESA, R., McKELLOP, H. A., DORR, L. D.: Accuracy of measurement of polyethylene wear with use of radiographs of total hip replacements. J. Bone Jt Surg., 85-A: 2378-2384, 2003. Přejít k původnímu zdroji... Přejít na PubMed...
  6. GALLO, J., HAVRÁNEK, V., HRABOVSKÝ, M.: Měření opotřebení extrahovaných polyetylenových jamek TEP kyčelního kloubu za pomoci univerzálního mikroskopu. Jemná mechanika a optika, 48: 333-338, 2003.
  7. GALLO, J., KAMÍNEK, P., ZAPLETALOVÁ, J., NOVOTNÝ, R., ČECHOVÁ, I., DITMAR, R.: Předčasné selhání polyetylenové vložky u TEP kyčelního kloubu ABG I. Acta Chir. orthop. Traum. čech, 70: 343-349, 2003.
  8. HENDL, J.: Přehled statistických metod zpracování dat. Analýza a metaanalýza dat. Praha, Portál 2004.
  9. JASTY, M., GOETZ, D. D., BRAGDON, C. R., LEE, K. R., HANSON, A. E., ELDER, J. R., HARRIS, W. H.: Wear of polyethylene acetabular components in total hip arthroplasty. An analysis of one hundred and twenty-eight components retrieved at autopsy or revision operations. J. Bone Jt Surg., 79-A: 349-358, 1997. Přejít k původnímu zdroji...
  10. KABO, J. M., GEBHARD, J. S., LOREN, G., AMSTUTZ, H. C.: In vivo wear of polyethylene acetabular components. J. Bone Jt Surg., 75-B: 254-258, 1993. Přejít k původnímu zdroji... Přejít na PubMed...
  11. KERRIGAN, S., RICKETTS, I. W., MCKENNA, S. J., WIGDEROWITZ, C.: Measurement of acetabular wear using intelligent ellipses. In: Meeting of the British Orthopaedic Research Society. Cardiff, p. 35, 2002.
  12. KNAHR, K., KARAMAT, L., LOHO, G., POSPISCHILL, M.: Tribological considerations for a new hip system. Acta Chir. orthop. Traum. čech., 72: 116-121, 2005.
  13. LEWIS, G.: Design issues in clinical studies of the in vivo volumetric wear rate of polyethylene bearing components. J. Bone Jt Surg., 82-A: 281-287, 2000. Přejít k původnímu zdroji... Přejít na PubMed...
  14. LIVERMORE, J., ILSTRUP, D., MORREY, B.: Effect of femoral head size on wear of the polyethylene acetabular component. J. Bone Jt Surg., 72-A: 518-528, 1990. Přejít k původnímu zdroji...
  15. MARTELL, J. M., BERDIA, S.: Determination of polyethylene wear in total hip replacements with use of digital radiographs. J. Bone Jt Surg., 79-A: 1635-1641, 1997. Přejít k původnímu zdroji... Přejít na PubMed...
  16. MARTELL, J. M., BERKSON, E., BERGER, R., JACOBS, J.: Comparison of two and three-dimensional computerized polyethylene wear analysis after total hip arthroplasty. J. Bone Jt Surg., 85-A: 1111-1117, 2003. Přejít k původnímu zdroji... Přejít na PubMed...
  17. McKELLOP, H. A., CAMPBELL, P., PARK, S. H., SCHMALZRIED, T. P., GRIGORIS, P., AMSTUTZ, H. C., SARMIENTO, A.: The origin of submicron polyethylene wear debris in total hip arthroplasty. Clin. Orthop., 311: 3-20, 1995.
  18. MIZOUE, T., YAMAMOTO, K., MASAOKA, T., IMAKIIRE, A., AKAGI, M., CLARKE, I. C.: Validation of acetabular cup wear volume based on direct and two-dimensional measurements: hip simulator analysis. J. Orthop. Sci., 8: 491-499, 2003. Přejít k původnímu zdroji... Přejít na PubMed...
  19. RAIMONDI, M. T., SASSI, R., PIETRABISSA, R.: A method for the evaluation of the change in volume of retrieved acetabular cups. Proc. Inst. Mech. Eng. [H], 214: 577-587, 2000. Přejít k původnímu zdroji... Přejít na PubMed...
  20. REDHEAD, A. L., KOTCHEFF, A. C., TAYLOR, C. J., PORTER, M. L., HUKINS, D. W.: An automated method for assessing routine radiographs of patients with total hip replacements. Proc. Inst. Mech. Eng. [H], 211: 145-154, 1997. Přejít k původnímu zdroji... Přejít na PubMed...
  21. SCHMALZRIED, T. P., CALLAGHAN, J. J.: Wear in total hip and knee replacements. J. Bone Jt Surg., 81-A: 115-136, 1999. Přejít k původnímu zdroji... Přejít na PubMed...
  22. SOSNA, A., RADONSKÝ, T., POKORNÝ, D., VEIGL, D., HORÁK, Z., JAHODA, D.: Polyetylenová nemoc. Acta Chir. orthop. Traum. čech., 70: 6-16, 2003.
  23. SYCHTERZ, C. J., MOON, K. H., HASHIMOTO, Y., TEREFENKO, K. M., ENGH, C. A., JR., BAUER, T. W.: Wear of polyethylene cups in total hip arthroplasty. A study of specimens retrieved post mortem. J. Bone Jt Surg., 78-A: 1193-1200, 1996. Přejít k původnímu zdroji... Přejít na PubMed...
  24. TIPPER, J. L., INGHAM, E., HAILEY, J. L., BESONG, A. A., FISHER, J., WROBLEWSKI, B. M., STONE, M. H.: Quantitative analysis of polyethylene wear debris, wear rate and head damage in retrieved Charnley hip prostheses. J. Mater. Sci. Mater. Med., 11: 117-124, 2000. Přejít k původnímu zdroji... Přejít na PubMed...
  25. YAMAGUCHI, M., HASHIMOTO, Y., AKISUE, T., BAUER, T. W.: Polyethylene wear vector in vivo: a three-dimensional analysis using retrieved acetabular components and radiographs. J. Orthop. Res., 17: 695-702, 1999. Přejít k původnímu zdroji... Přejít na PubMed...