Acta Chir Orthop Traumatol Cech. 2013; 80(4):273-277 | DOI: 10.55095/achot2013/045

Accuracy of Proximal Femur Correction Achieved with LCP Paediatric Hip PlatesOriginal papers

K. Ročák1, J. Poul2,*, K. Urbášek2
1 Oddělení ortopedie a traumatologie, o.z Nemocnice Přerov, Středomoravská nemocniční a.s, člen skupiny AGEL
2 Klinika dětské chirurgie, ortopedie a traumatologie, Fakultní nemocnice Brno

PURPOSE OF THE STUDY:
The aim of the study was to evaluate, in comparison with the pre-operative planning, the accuracy of proximal femur correction achieved with the use of locking compression paediatric hip plates (LCP) in children and adolescents and to assess pre- and post-operative complications.

MATERIAL AND METHODS:
A group of 52 patients in whom proximal femoral osteotomy using the LCP was performed on a total of 55 hips between September 2009 and February 2013 were retrospectively evaluated. The following diagnoses were treated: unstable hip in cerebral palsy, 18 operations; Legg-Calvé-Perthes disease, 10 operations; coxa vara of aetiology other than coxa vara adolescentium (CVA), eight operations; true CVA, six operations; femoral shortening by the Wagner method, six procedures; proximal femoral derotation osteotomy, four procedures; and post-traumatic pseudoarthrosis of the proximal femur, three operations.

RESULTS:
Compared with the pre-operative plan, the average deviation of the colodiaphyseal angle was 5.2° (1° to 11°) in 18 unstable hips; 4.7° (1° to 10°) in 10 cases of Legg-Calvé-Perthes disease; 4.5° (3° to 6°) in eight hips with coxa vara of aetiology other than CVA; 6.5° (2° to 13°) in six CVA hips; 4.5° (1° to 10°) in six cases of femoral shortening; 3.5° (1° to 5°) in four derotation osteotomies; and 3.7° (0° to 6°) in three corrections of pseudoarthrosis. In one patient, osteosynthesis failed due to screws being pulled out from the proximal fragment; re-osteosynthesis was carried out using a conventional angled blade plate.

DISCUSSION:
As in other international studies, our results confirmed a high accuracy of proximal femur correction with use of the LCP instrumentation. The reported higher time requirement for this technique seems to be related to the learning curve and, with more frequent use, will probably be comparable to the time needed for application of conventional hip angled plates.

CONCLUSIONS:
The up-to-date LCP fixation system using the principle of angular stability for correction of the proximal femur in children is a clear advancement and its higher costs are certain to be outweighed by its higher accuracy and thus better results.

Keywords: locking compression plate, paediatric hip

Published: August 1, 2013  Show citation

ACS AIP APA ASA Harvard Chicago Chicago Notes IEEE ISO690 MLA NLM Turabian Vancouver
Ročák K, Poul J, Urbášek K. Accuracy of Proximal Femur Correction Achieved with LCP Paediatric Hip Plates. Acta Chir Orthop Traumatol Cech. 2013;80(4):273-277. doi: 10.55095/achot2013/045. PubMed PMID: 24119475.
Download citation

References

  1. BECKER, T., WEIGL, D., MERCADO, E., KATZ, K., BAR-ON, E.: Fractures and refractures after femoral locking plate fixation in children and adolescents. J. Pediatr. Orthop., 32: 40-46, 2012. Go to original source... Go to PubMed...
  2. ČECH, O., STRYHAL, F.: Moderní osteosyntéza v traumatologii a ortopedii. Praha, Avicenum 1972.
  3. ČECH, O., STRYHAL, F., SOSNA, A., BEZNOSKA, S.: Stabilní osteosyntéza v traumatologii a ortopedii. Praha, Avicenum 1982.
  4. ČECH, O., VÁVRA, J.: Valgizující osteotomie při léčení následků ischemické postdysplastické nekrózy hlavice stehenní kosti u dětí. Acta Chir. orthop. Traum. čech., 65: 69-73, 1998.
  5. ČECH, O. VÁVRA, J., ZÍDKA, M.: Management of ischemic deformity after the treatment of developmental dysplasia of the hip. J. Pediatr. Orthop., 25: 687-694, 2005 Go to original source... Go to PubMed...
  6. deMORAIS FILHO, M. C., NEVES, D. L., ABREU, F. P., KAWAMURA, C. M., DOS SANTOS, C. A.: Does the level of proximal femur rotation osteotomy influence the correction results in patients with cerebral palsy? J. Pediatr. Orthop. B., 2013 Jan. 22 : 8-13 (Epub). Go to original source... Go to PubMed...
  7. DOSTÁL, M., DITMAR, R., PACH, M.: První zkušenosti s Morscherovou elongační osteotomií u coxa brevis. Acta Chir. orthop. Traum. čech., 61: 171-176, 1994.
  8. DUNGL, P.: Patnáctileté výsledky operačního léčení Perthesova onemocnění. Acta Chir. orthop. Traum. čech., 63: 269-283, 1996
  9. FRIGG, R.: Locking Compression Plate (LCP). An osteosynthesis plate based on the dynamic compression plate and the Point Contact Fixator (PC-Fix) Injury, 32 (Suppl. 2): 63-66, 2001. Go to original source... Go to PubMed...
  10. FUJAK, A., MÜLLER, K., LEGAL, W.,, LEGAL, H., FORST, R., FORST, J.: Long-term results of Imhäuser osteotomy for chronic slipped femoral head epiphyseolysis. Orthopäde, 41: 452-458, 2012. Go to original source... Go to PubMed...
  11. GAUTIER, E., SOMMER, C.: Guidelines for the clinical application of the LCP. Injury, 34 (Suppl. 2): B63-76, 2003. Go to original source... Go to PubMed...
  12. HAEFELI, M., HUBER, H., DIERAUER, S., RAMSEIER, L. E.: Fixation of subtrochanteric extending/derotational femoral osteotomies with the locking compression plate in ambulatory neuro-orthopaedic patients. J. Child. Orthop., 4: 423-428, 2010. Go to original source... Go to PubMed...
  13. HOERDEMANN, M., GÉDET, P., FERGUSON, S. J., SAUTER-LOUIS, C., NUSS, K.: In-vitro comparison of LC-DCP - and LCP-constructs in the femur of newborn calves-a pilot study. BMC Vet. Res., 8: 139, 2012. Go to original source... Go to PubMed...
  14. HUBER, H., HAEFELI, M., DIERAUER, S., RAMSEIER, L. E.: Treatment of reduced femoral antetorsion by subtrochanteric rotational osteotomy. Acta Orthop. Belg., 75: 490-496, 2009.
  15. JOERIS, A., AUDIGÉ, L., ZIEBARTH, K., SLONGO, T.: The locking compression paediatric hip plate: technical quide and critical analysis. Int. Orthop., 2012 Aug. 26 (Epub). Go to original source... Go to PubMed...
  16. KHOURI, N., KHALIFE, R., DESAILLY, E., THEVENIN-LEMOINE, C., DAMSIN, J. P.: Proximal femoral osteotomy in neurologic pediatric hips using the locking compression plate. J. Pediatr. Orthop., 30: 825-831, 2010. Go to original source... Go to PubMed...
  17. KIM, J. W., OH C. W., BYUN, Y. S., OH J. K., KIM, H. J., MIN, W. K., PARK, S. K., PARK, B. C.: A biomechanical analysis of locking plate fixation with minimally invasive plate osteosynthesis in a subtrochanteric fracture model. J. Trauma, 70: 19-23, 2011. Go to original source... Go to PubMed...
  18. ROZKYDAL, Z., JANÍČEK, P., OTIEPKA, P.: Varizační osteotomie proximálního konce femuru u dospělých po vývojové dysplázii kyčle-dlouhodobé výsledky. Acta Chir. orthop. Traum. čech., 77: 489-493, 2010. Go to original source...
  19. RUTZ, E., BRUNNER, R.: The pediatric LCP hip plate for fixation of proximal femoral osteotomy in cerebral palsy and severe osteoporosis. J. Pediatr. Orthop., 7: 726-731, 2010. Go to original source... Go to PubMed...
  20. SCHEJBALOVÁ, A.: Derotační subtrochanterická osteotomie femuru u pacientů s dětskou mozkovou obrnou. Acta Chir. orthop. Traum. čech., 73: 334-339, 2006. Go to original source...
  21. SCHÜTZ, M., SÜDKAMP, N. P.: Revolution in plate osteosynthesis: new internal fixator systém. J. Orthop. Sci., 8: 252-258, 2003. Go to original source... Go to PubMed...
  22. SMETANA, V., CHLÁDEK, P.: Ischemická postluxační nekróza hlavice femuru: možnosti chirurgického léčení. I. část: teoretický úvod, konzervativní a chirurgická léčba. Acta Chir. orthop. Traum. čech., 61: 218-223, 1994
  23. SMETANA, V., CHLÁDEK, P.: Ischemická postluxační nekróza hlavice femuru: možnosti chirurgického léčení. II. část: operační postup na ortopedické klinice FN Motol. Acta Chir. orthop. Traum. čech., 61: 271-275, 1994
  24. STARÝ, D., POUL, J., CHARVÁTOVÁ, M., GÁL, P.: Osteotomie proximálního femuru kanalizovaným pediatrickým ortopedickým systémem. Ortopedie, 3: 56-59, 2009.
  25. VÁVRA, J, ČECH, O., ZÍDKA, M.: Valgizující osteotomie při léčení následků ischemické postdysplastické nekrózy hlavice stehenní kosti u dětí-II. část: klinické zhodnocení souboru. Acta Chir. orthop. Traum. čech., 67: 77-87, 2000.
  26. WAGNER, M., FRENK, A., FRIGG, R.: New concepts for bone fracture treatment and the Locking Compression Plate. Surg. Technol. Int., 12: 271-277, 2004.