Acta Chir Orthop Traumatol Cech. 2015; 82(3):235-238 | DOI: 10.55095/achot2015/035

Ohybová tuhost zubu čepovce po přední fixaci jedním tahovým šroubem - biomechanická studiePůvodní práce

P. BUCHVALD1,*, L. ČAPEK2, P. BARSA1
1 Neurocentrum, Krajská nemocnice Liberec, a. s.
2 Fakulta strojní, Technická univerzita Liberec

PURPOSE OF THE STUDY:
The aim of the experiment was to compare the bending stiffness of an intact odontoid process with bending stiffness after its simulated type II fracture was fixed with a single lag screw. The experiment was done with a desire to answer the question of whether a single osteosynthetic screw is sufficient for good fixation of a type II odontoid fracture.

MATERIAL AND METHODS:
The C2 vertebrae of six cadavers were used. With simultaneous measurement of odontoid bending stiffness, the occurrence of a fracture (type IIA, Grauer's modification of the Anderson- D'Alonzo classification) was simulated using action exerted by a tearing machine in the direction perpendicular to the odontoid axis. Each odontoid fracture was subsequently treated by direct osteosynthesis with a single lag screw inserted in the axial direction by a standard surgical procedure in order to provide conditions similar to those achieved by routine surgical management. The treated odontoid process was subsequently subjected to the same tearing machine loading as applied to it at the start of the experiment. The bending stiffness measured was then compared with that found before the fracture occurred. The results were statistically evaluated by the t-test for paired samples at the level of significance α = 0.05.

RESULTS:
The average value of bending stiffness for odontoid processes of intact vertebrae at the moment of fracture occurrence was 318.3 N/mm. After single axial lag screw fixation of the fracture, the average bending stiffness for the odontoid processes treated was 331.3 N/mm.

DISCUSSION:
Higher values of bending stiffness after screw fixation were found in all specimens and, in comparison with the values recorded before simulated fractures, the increase was statistically significant.

CONCLUSIONS:
The results of our measurements suggest that the single lag screw fixation of a type IIA odontoid fracture will provide better stability for the fracture fragment-C2 body complex on antero-posterior perpendicular loading than can be found in intact C2 vertebrae.

Klíčová slova: odontoid fracture, odontoid fixation, bending stiffness, lag screw

Zveřejněno: 1. červen 2015  Zobrazit citaci

ACS AIP APA ASA Harvard Chicago Chicago Notes IEEE ISO690 MLA NLM Turabian Vancouver
BUCHVALD P, ČAPEK L, BARSA P. Ohybová tuhost zubu čepovce po přední fixaci jedním tahovým šroubem - biomechanická studie. Acta Chir Orthop Traumatol Cech. 2015;82(3):235-238. doi: 10.55095/achot2015/035. PubMed PMID: 26317296.
Stáhnout citaci

Reference

  1. ALFIERI, A.: Single-screw fixation for acute type II odontoid fracture. J. Neurosurg. Sci., 45: 15-18, 2001.
  2. ANDERSON, L. D., D'ALONZO, R. T.: Fractures of the odontoid process of the axis. J. Bone Jt Surg., 56-A: 1663-1674, 1974. Přejít k původnímu zdroji...
  3. BOEHLER, J.: Anterior stabilization for acute fractures and nonunions of the dens. J. Bone Joint Surg., 64-A: 18-27, 1982. Přejít k původnímu zdroji...
  4. BOEHLER, J.: Schraubenosteosynthese von fracturen des dens axis. Umfallheilkunde, 84: 221-223, 1981.
  5. DOHERTY, B. J., HEGGENESS, M. H., ESSES, S. I.: A biomechanical study of odontoid fractures and fracture fixation. Spine, 18: 178-184, 1993. Přejít k původnímu zdroji... Přejít na PubMed...
  6. FAN, K. F., LIAO, J. C., NIU, C. C., CHEN, L. H., CHEN, W. J, LEE, M. S.: Anterior single- screw fixation in 24 patients with type II odontoid fractures. F. J. Muscul. Disord. 4: 26-31, 2013. Přejít k původnímu zdroji...
  7. GRAUER, J. N., SHAFI, B., HILIBRAND, A. S.: Proposal of a modified, treatment - oriented classification of odontoid fractures. Spine J., 5: 123-129, 2005. Přejít k původnímu zdroji... Přejít na PubMed...
  8. GRAZIANO, G., JAGGERS, C., LEE, M., LYNCH, W.: A comparative study of fixation techniques for type II fractures of the odontoid process. Spine, 18: 2383-2387, 1993. Přejít k původnímu zdroji... Přejít na PubMed...
  9. HOU, Y., YUAN, W., WANG, X.: Clinical evaluation of anterior screw fixation for elderly patients with type II. odontoid fractures. J. Spinal. Disord. Tech., 24: 75-81, 2011. Přejít k původnímu zdroji... Přejít na PubMed...
  10. HRABÁLEK, L., BUŘVAL, S., VAVERKA, M.: Úspěšnost přední osteosyntézy fraktury dens axis. Acta Chir. orthop. Traum. čech, 75: 332-338, 2008. Přejít k původnímu zdroji...
  11. JENKINS, J. D., CORIC, D., BRANCH, C. L. Jr.: A clinical comparison of one- and two-screw odontoid fixation. J. Neurosurg., 89: 366-370, 1998. Přejít k původnímu zdroji... Přejít na PubMed...
  12. LEE, S. C., CHEN, J.F., LEE, S. T.: Management of acute odontoid fractures with single anterior screw fixation. J. Clin. Neurosci., 11: 890-895, 2004. Přejít k původnímu zdroji... Přejít na PubMed...
  13. MOURADIAN, W., FIETTI, V. J., COCHRAN, G., FIELDING, J., YOUNG, J.: Fractures of the odontoid: a laboratory and clinical study of mechanisms. Orthop. Clin. North. Am., 9: 985-100, 1978. Přejít k původnímu zdroji...
  14. NAKANISHI, T.: Internal fixation of the odontoid fracture. Cent. J. Orthop. Trauma Surg., 23: 399-406, 1980.
  15. RYAN, M. D., HENDERSON, J. J.: The epidemiology of fractures and fracture - dislocations of the cervical spine. Injury, 23: 38-40, 1992. Přejít k původnímu zdroji... Přejít na PubMed...
  16. SASSO, R., DOHERTY, B. J., CRAWFORD, M. J., HEGGENESS, M. H.: Biomechanics of odontoid fracture fixation: comparison of the one- and two-screw technique. Spine, 18: 1950-1953, 1993. Přejít k původnímu zdroji... Přejít na PubMed...
  17. SAUR, K., SAMEŠ, M.: Výsledky ošetření zlomenin zubu čepovce přímou osteosyntézou jediným axiálním šroubem. Acta Chir. orthop. Traum. čech., 75: 48-51, 2008. Přejít k původnímu zdroji...
  18. SAYED AHMED, N. M., LOUTFY, M. I., SHERSHERA, W., SLEEM, A.: Fixation of type II odontoid fractures with anterior single screw. E. J. N. S. 22: 137-146, 2007.
  19. SUBACH, B. R., MORONE, M. A., HAID, R. W. JR., McLAUGHLIN, M. R., RODTS, G. R., COMEY, C. H.: Management of acute odontoid fractures with single-screw anterior fixation. Neurosurgery, 45: 812-819; discussion 819-820, 1999. Přejít k původnímu zdroji... Přejít na PubMed...
  20. ŠTULÍK, J., SUCHOMEL, P., LUKÁŠ, R., CHROBOK, J., KLÉZL, Z., TALLER, S., KRBEC, M.: Přímá osteosynthesa dentu - multicentrická studie. Acta Chir. orthop. Traum. čech., 69: 141-148, 2002.