Acta Chir Orthop Traumatol Cech. 2016; 83(5):344-347 | DOI: 10.55095/achot2016/054

Surgical Correction of Scoliosis: Does Intraoperative CT Navigation Prolong Operative Time?Original papers

J. SKÁLA-ROSENBAUM1,*, M. JEŽEK1, V. DŽUPA1,2, R. KADEŘÁBEK3, P. DOUŠA1, R. RUSNÁK2,4, M. KRBEC1
1 Ortopedicko-traumatologická klinika, 3. lékařská fakulta Univerzity Karlovy a Fakultní nemocnice Královské Vinohrady, Praha
2 Centrum pro integrované studium pánve, 3. lékařská fakulta Univerzity Karlovy, Praha
3 Radiodiagnostická klinika, 3. lékařská fakulta Univerzity Karlovy a Fakultní nemocnice Královské Vinohrady, Praha
4 Neurochirurgická klinika, Ústredná vojenská nemocnica SNP a FN, Ružomberok

PURPOSE OF THE STUDY:
The aim of the study was to compare the duration of corrective surgery for scoliosis in relation to the intra-operative use of either fluoroscopic or CT navigation.

MATERIAL AND METHODS:
The indication for surgery was adolescent idiopathic scoliosis in younger patients and degenerative scoliosis in middleage or elderly patients. In a retrospective study, treatment outcomes in 43 consecutive patients operated on between April 2011 and April 2014 were compared. Only patients undergoing surgical correction of five or more spinal segments (fixation of six and more vertebrae) were included.

RESULTS:
Transpedicular screw fixation of six to 13 vertebrae was performed under C-arm fluoroscopy guidance in 22 patients, and transpedicular screws were inserted in six to 14 vertebrae using the O-arm imaging system in 21 patients. A total of 246 screws were placed using the C-arm system and 340 screws were inserted using the O-arm system (p < 0.001). The procedures with use of the O-arm system were more complicated and required an average operative time longer by 48% (measured from the first skin incision to the completion of skin suture). However, the mean time needed for one screw placement (the sum of all surgical procedures with the use of a navigation technique divided by the number of screws placed using this technique) was the same in both techniques (19 min).

DISCUSSION:
With good teamwork (surgeons, anaesthesiologists and a radiologist attending to the O-arm system), the time required to obtain one intra-operative CT scan is 3 to 5 minutes. The study showed that the mean time for placement of one screw was identical in both techniques although the average operative time was longer in surgery with O-arm navigation. The 19-minute interval was not the real placement time per screw. It was the sum of all operative times of surgical procedures (from first incision to suture completion including the whole approach within the range of planned stabilization) which used the same navigation technique divided by the number of all screws inserted during the procedures. The longer average operative time in procedures using O-arm navigation was not related to taking intra-operative O-arm scans. The authors consider surgery with an O-arm imaging system to be a safer procedure and use it currently in surgical correction of scoliosis.

CONCLUSIONS:
The study focused on the length of surgery to correct scoliosis performed using either conventional fluoroscopy (C-arm) or intra-operative CT scanning (O-arm) showed that the mean placement time for one screw was identical in both imaging techniques when six or more vertebrae were stabilised. The use of intra-operative CT navigation did not make the surgery longer, and the higher number of inserted screws provides evidence that this technique is safer and allows us to achieve good stability of the correction procedure.

Keywords: virtual CT guidance, O-arm, scoliosis, transpedicular screw

Published: October 1, 2016  Show citation

ACS AIP APA ASA Harvard Chicago Chicago Notes IEEE ISO690 MLA NLM Turabian Vancouver
SKÁLA-ROSENBAUM J, JEŽEK M, DŽUPA V, KADEŘÁBEK R, DOUŠA P, RUSNÁK R, KRBEC M. Surgical Correction of Scoliosis: Does Intraoperative CT Navigation Prolong Operative Time? Acta Chir Orthop Traumatol Cech. 2016;83(5):344-347. doi: 10.55095/achot2016/054. PubMed PMID: 28102810.
Download citation

References

  1. Burch S, Karahalios D, Mobasser J-P, Potts E. Comparison of radiation exposure to the spine surgeon during pedicle screw placement using O-arm® System & StealthStation® Navigation vs C-arm Standard Flouroscopy. Medtronic, Louisville, 2010, pp.1-5.
  2. Džupa V, Krbec M, Kadeřábek R, Rusnák R, Douša P, Skála-Rosenbaum J, Fridrich F, Báča V, Grill R. Peroperační CT navigace u operací páteře a pánve: první zkušenosti. Rozhl Chir. 2013; 92:379-384. Go to PubMed...
  3. Kahler DM. Computer-assisted closed techniques of reduction and fixation In: Tile M, Helfet DL, Kellam JE (Eds). Fractures of the pelvis and acetabulum. Third edition, Lippincott Williams & Wilkins, Philadelphia, 2003, pp.604-615.
  4. Kendoff D, Citak M, Hüfner T, Chaudhary S, Krettek C. Current concepts and applications of computer navigation in orthopedic trauma surgery. CEJMed. 2007;2:392-403. Go to original source...
  5. Kendoff D, Gardner MJ, Citak M, Kfuri M, Thumes B, Krettek C, Hüfner T. Value of 3D fluoroscopic imaging of acetabular fractures comparison to 2D fluoroscopy and CT imaging. Arch Orthop Trauma Surg. 2008;128:599-605. Go to original source... Go to PubMed...
  6. Mosheiff R, Khoury A, Weil Y, Liebergall M. First generation computerized fluoroscopic navigation in percutaneous pelvic surgery. J Orthop Trauma. 2004;18:106-111. Go to original source... Go to PubMed...
  7. Ochs BG, Gonser C, Shiozawa T, Badke A, Weise K, Rolauffs B, Stuby FM. Computer-assisted periacetabular screw placement:comparison of different fluoroscopy-based navigation procedures with conventional technique. Injury. 2010;41:1297-1305. Go to original source... Go to PubMed...
  8. Park MS, Chung CY, Kwon DG, Huh Y, Lee K, Lee KM. The perspectives of user and developers in designing and developing O-arm Imaging System. J X-ray Science Technol. 2011;19:199-204. Go to original source... Go to PubMed...
  9. Patil S, Lindley Em, Burger EL, Yoshihara H, Patel VV. Pedicle screw placement with O-arm and stealth navigation. Orthopedics. 2012;35:e61-65. Go to original source... Go to PubMed...
  10. Stöckle U, Schäffler A, König B, Haas NP. Integrierte Navigation: präklinische Erprobung und erste klinische Erfahrungen. Unfallchirurg. 2006;109:925-931. Go to original source... Go to PubMed...
  11. Van de Kelft E, Vosta F, Van der Planken D, Schils F. A prospective multicenter registry on the accuracy of pedicle screw placement in the thoracic lumbar and sacral levels with the use of the O-arm Imaging System and StealthStation Navigation Spine (Phila Pa 1976). 2012;37:E1580-1587. Go to original source... Go to PubMed...
  12. Zhang J, Weir V, Fajardo L, Lin J, Hsiung H, Ritenour ER. Dosimetric characterization of a cone-beam O-arm (TM) Imaging System. J X-ray Sci Technol. 2009;17:305-317. Go to original source... Go to PubMed...