Acta Chir Orthop Traumatol Cech. 2015; 82(5):341-347 | DOI: 10.55095/achot2015/057
Determination of a Friction Coefficient for THA Bearing CouplesOriginal papers
- 1 Ústav konstruování, Fakulta strojního inľenýrství, VUT v Brně, Brno
- 2 Ústav lékařské chemie a biochemie LF UP v Olomouci, Olomouc
- 3 Ortopedická klinika LF UP a FN Olomouc, Olomouc
PURPOSE OF THE STUDY:
The wear of articular surfaces is considered one of the most important factors limiting the life of total hip arthroplasty (THA). It is assumed that the particles released from the surface of a softer material induce a complex inflammatory response, which will eventually result in osteolysis and aseptic loosening. Implant wear is related to a friction coefficient which depends on combination of the materials used, roughness of the articulating surfaces, internal clearance, and dimensions of the prosthesis.
MATERIAL AND METHODS:
The selected parameters of the bearing couples tested were studied using an experimental device based on the principle of a pendulum. Bovine serum was used as a lubricant and the load corresponded to a human body mass of 75 kg. The friction coefficient was derived from a curve of slowdown of pendulum oscillations. Roughness was measured with a device working on the principle of interferometry. Clearance was assessed by measuring diameters of the acetabular and femoral heads with a 3D optical scanner. The specimens tested included unused metal-on-highly cross-linked polyethylene, ceramic-on-highly cross-linked polyethylene and ceramic-on-ceramic bearing couples with the diameters of 28 mm and 36 mm. For each measured parameter, an arithmetic mean was calculated from 10 measurements.
RESULTS:
1) The roughness of polyethylene surfaces was higher by about one order of magnitude than the roughness of metal and ceramic components. The Protasul metal head had the least rough surface (0.003 μm).
2) The ceramic-on-ceramic couples had the lowest clearance. Bearing couples with polyethylene acetabular liners had markedly higher clearances ranging from 150 μm to 545 μm. A clearance increased with large femoral heads (up to 4-fold in one of the couple tested).
3) The friction coefficient was related to the combination of materials; it was lowest in ceramic-on-ceramic surfaces (0.11 to 0.12) and then in ceramic-on-polyethylene implants (0.13 to 0.14). The friction coefficient is supposed to increase with a decreasing femoral head diameter. However, in the bearing couples with polyethylene liners manufactured by one company, paradoxically, the friction coefficient slightly increased with an increase in femoral head size from 28 mm to 36 mm.
4) The lowest friction moment (< 3.5 Nm) was found for ceramic-on-ceramic implants 28 mm in diameter; the highest values were recorded in metal-on-polyethylene bearing couples 36 mm in diameter (> 7 Nm).
DISCUSSION:
Although our study confirmed that the bearing couples produced by different manufacturers varied to some extent in the parameters studied, in our opinion, this variability was not significant because it was not within an order of magnitude in any of the tests.
CONCLUSIONS:
The study showed that both the friction coefficient and the friction moment are affected more by the combination of materials than by the diameter of a femoral head. The best results were achieved in ceramic-on-ceramic implants.
Keywords: total hip arthroplasty, bearing surfaces, wear, friction coefficient, friction moment, clearance
Published: October 1, 2015 Show citation
ACS | AIP | APA | ASA | Harvard | Chicago | Chicago Notes | IEEE | ISO690 | MLA | NLM | Turabian | Vancouver |
References
- AL-HAJJAR, M., FISHER, J., TIPPER, J. L., WILLIAMS, S., JENNINGS, L. M.: Wear of 36-mm BIOLOX(R) delta ceramic-on-ceramic bearing in total hip replacements under edge loading conditions. Proc. Inst. Mech. Eng. Part H, 227: 535-542, 2013.
Go to original source...
Go to PubMed...
- BISHOP, N. E., WALDOW, F., MORLOCK, M. M.: Friction moments of large metal-on-metal hip joint bearings and other modern designs. Med. Eng. Phys., 30: 1057-1064, 2008.
Go to original source...
Go to PubMed...
- BRAGDON, C. R., DOERNER, M., MARTELL, J., JARRETT, B., PALM, H., Multicenter Study Group, MALCHAU, H.: The 2012 John Charnley Award: Clinical multicenter studies of the wear performance of highly crosslinked remelted polyethylene in THA. Clin. Orthop. Relat. Res., 471: 393-402, 2013.
Go to original source...
Go to PubMed...
- BROCKETT, C. L., HARPER, P., WILLIAMS, S., ISAAC, G. H., DWYER-JOYCE, R. S., JIN, Z., FISHER, J.: The influence of clearance on friction, lubrication and squeaking in large diameter metal-on-metal hip replacements. J. Mater. Sci. Mater. Med., 19: 1575-1579, 2008.
Go to original source...
Go to PubMed...
- CALLARY, S. A., SOLOMON, L. B., HOLUBOWYCZ, O. T., CAMPBELL, D. G., MUNN, Z., HOWIE, D. W.: Wear of highly crosslinked polyethylene acetabular components. Acta Orthop.: 1-10, 2014.
Go to original source...
Go to PubMed...
- CRISCO, J. J., BLUME, J., TEEPLE, E., FLEMING, B. C., JAY, G. D.: Assuming exponential decay by incorporating viscous damping improves the prediction of the coefficient of friction in pendulum tests of whole articular joints. Proc. Inst. Mech. Eng. H, 221: 325-333, 2007.
Go to original source...
Go to PubMed...
- DAMM, P., DYMKE, J., ACKERMANN, R., BENDER, A., GRAICHEN, F., HALDER, A., BEIER, A., BERGMANN, G.: Friction in total hip joint prosthesis measured in vivo during walking. PLoS One, 8: e78373, 2013.
Go to original source...
Go to PubMed...
- DOWSON, D.: Bio-tribology. Faraday Discuss, 156: 9-30, 2012.
Go to original source...
Go to PubMed...
- DOWSON, D., HARDAKER, C., FLETT, M., ISAAC, G. H.: A hip joint simulator study of the performance of metal-on-metal joints: Part II: design. J. Arthroplasty, 19: 124-130, 2004.
Go to original source...
Go to PubMed...
- FULÍN, P., POKORNÝ, D., ©LOUF, M., VACKOVÁ, T., DYBAL, J., SOSNA, A.: Vliv sterilizace formaldehydem, gama zářením a etylenoxidem na vlastnosti polyetylenových komponent kloubních náhrad. Acta Chir. orthop. Traum. čech., 81: 33-39, 2014.
Go to original source...
- GALLO, J., GOODMAN, S. B., KONTTINEN, Y. T., RASKA, M.: Particle disease: biologic mechanisms of periprosthetic osteolysis in total hip arthroplasty. Innate Immun., 19: 213-224, 2013.
Go to original source...
Go to PubMed...
- GALLO, J., GOODMAN, S. B., LOSTAK, J., JANOUT, M.: Advantages and disadvantages of ceramic on ceramic total hip arthroplasty: a review. Biomed. Pap. Med. Fac. Univ. Palacky Olomouc Czech Repub., 156: 204-212, 2012.
Go to original source...
Go to PubMed...
- GALLO, J., HARTL, M., VRBKA, M., NAVRAT, T., KRUPKA, I.: Co by měl vědět ortoped o biotribologii totální endoprotézy kyčelního kloubu. Acta Chir. orthop. Traum. čech., 80: 377-385, 2013.
Go to original source...
- GALLO, J., LANGOVA, K., HAVRANEK, V., CECHOVA, I.: Poor survival of ABG I hip prosthesis in younger patients. Biomed. Pap. Med. Fac. Univ. Palacky Olomouc Czech Repub., 152: 163-168, 2008.
Go to original source...
Go to PubMed...
- CHARNLEY, J.: Total hip replacement by low-friction arthroplasty. Clin. Orthop. Relat. Res., 72: 7-21, 1970.
Go to original source...
- JAMESON, S. S., LEES, D., JAMES, P., SERRANO-PEDRAZA, I., PARTINGTON, P. F., MULLER, S. D., MEEK, R. M., REED, M. R.: Lower rates of dislocation with increased femoral head size after primary total hip replacement: a five-year analysis of NHS patients in England. J. Bone Jt Surg., 93-B: 876-880, 2011.
Go to original source...
Go to PubMed...
- JIN, Z. M., STONE, M., INGHAM, E., FISHER, J.: Biotribology. Current Orthopaedics, 20: 32-40, 2006.
Go to original source...
- LEHIL, M. S., BOZIC, K. J.: Trends in total hip arthroplasty implant utilization in the United States. J. Arthroplasty, 29: 1915-1918, 2014.
Go to original source...
Go to PubMed...
- MA, S. M., KABO, J. M., AMSTUTZ, H. C.: Frictional torque in surface and conventional hip replacement. J. Bone Jt. Surg., 65-A: 366-370, 1983.
Go to original source...
- MASSIN, P., LOPES, R., MASSON, B., MAINARD, D., FRENCH, H., KNEE, S.: Does Biolox Delta ceramic reduce the rate of component fractures in total hip replacement? Orthop. Traumatol. Surg. Res., 100: S317-321, 2014.
Go to original source...
Go to PubMed...
- MATTEI, L., DI PUCCIO, F., PICCIGALLO, B., CIULLI, E.: Lubrication and wear modelling of artificial hip joints: A review. Tribology International, 44: 532-549, 2011.
Go to original source...
- OWEN, D. H., RUSSELL, N. C., SMITH, P. N., WALTER, W. L.: An estimation of the incidence of squeaking and revision surgery for squeaking in ceramic-on-ceramic total hip replacement: a meta-analysis and report from the Australian Orthopaedic Association National Joint Registry. Bone Joint J., 96-B: 181-187, 2014.
Go to original source...
Go to PubMed...
- SARIALI, E., STEWART, T., JIN, Z., FISHER, J.: In vitro investigation of friction under edge-loading conditions for ceramic-on-ceramic total hip prosthesis. J. Orthop. Res., 28: 979-985, 2010.
Go to original source...
Go to PubMed...
- STANTON, T. E.: Boundary lubrication in engineering practice. Engineer, 135: 678-680, 1923.
- SU, E. P.: Ceramic-ceramic bearing: Too unpredictable to use it regularly. HSS J., 8: 287-290, 2012.
Go to original source...
Go to PubMed...
- VRBKA, M., KRUPKA, I., HARTL, M., NAVRAT, T., GALLO, J., GALANDAKOVA, A.: In situ measurements of thin films in bovine serum lubricated contacts using optical interferometry. Proc. Inst. Mech. Eng. H, 228: 149-158, 2014.
Go to original source...
Go to PubMed...
- VRBKA, M., NAVRAT, T., KRUPKA, I., HARTL, M., SPERKA, P., GALLO, J.: Study of film formation in bovine serum lubricated contacts under rolling/sliding conditions. Proc. Inst. Mech. Eng. J, 227: 459-475, 2013.
Go to original source...