Wear and Roughness of Bearing Surface in Retrieved Polyethylene Bicon-Plus CupsOriginal papers

M. RANUŠA1, J. GALLO2,*, M. HOBZA2, M. VRBKA1, D. NEČAS1, M. HARTL1
1 Ústav konstruování, Fakulta strojního inženýrství, Vysoké učení technické v Brně
2 Ortopedická klinika, Lékařská fakulta, Univerzita Palackého v Olomouci a Fakultní nemocnice Olomouc

PURPOSE OF THE STUDY:
By 7th December 2016, 4,755 Bicon-Plus cups in total were implanted in the Czech Republic. Some of them have been continuously re-operated, while the most frequent reason of failure is polyethylene wear and aseptic loosening. The present study is focused on surface analysis of retrieved polyethylene Bicon-Plus cups and the determination of the roughness of their bearing surfaces.

MATERIAL AND METHODS:
In this study, we had 13 high molecular weight polyethylene cups with the average time in situ of 8.11 years (3.6-13.7, SD 3.2) before the retrieval. The study population was composed of 3 men, 10 women, with the mean age of 53.31 years. An optical scanning method, based on the principle of active triangulation, was used to determine wear rate. The rate of wear was identified by means of an obtained scan subsequently processed with the use of the GOM Inspect software. The roughness of surfaces was analysed with the application of Contour GT-X8 profiler using the principle of phase shifting interferometry. Measurements of surface topography of the retrieved cups were performed on the entire bearing surfaces. For the individual surface changes, a typical range of surface roughness, describing the particular wear character, was determined. By means of morphology analysis of the tested implants, three areas were identified: unworn area; area representing the worn part of the cup; and the area roughened by parallel grooving. The total surface roughness was evaluated as an arithmetic mean of the measured values. Subsequently, the values were sorted based on frequency and were classified into categories defining the particular wear mechanisms.

RESULTS:
Wear rate of the retrieved acetabular cups was evaluated based on the wear direction vector and the size of linear wear. The average linear wear was equal to 0.13 mm/year (ranging from 0.26 to 2.29 mm/year), and the mean value of total volumetric material loss was 44.37 mm3/year (the range being from 51.80 to 1,119.7 mm3/year). Using the optical profilometer, a map of roughness distribution of the individual cups was obtained. For each implant, 76 values of roughness were evaluated. With the respect to average roughness, the samples were sorted to various categories describing: surface polishing; abrasive-adhesive wear; surfaces with preserved grooving; substantial plastic deformation.

DISCUSSION:
The results clearly showed an increase of wear depending on implant survival; however, the tendency is not linear. This fact can be attributed to a larger amount of abrasive particles, causing an increase of wear or occurrence of surface wear in terms of micro cracks and oxidation degradation of polyethylene. This study indicates that geometry, positioning, and cup alignment during the implantation have a fundamental impact on the cup durability. Further correlation, which was observed in the case of the cup with prevailing roughness in the range from 0.1 to 0.3 μm, is a relatively wide wear vector angle determined with the use of the optical method. Considering the implants with the longest survival time with no loosening of the acetabular cup, the mean angle of direction vector was 56.8° (SD 2.1°).

CONCLUSIONS:
The present study provides the results of morphology analysis of the retrieved Bicon-Plus cups. In general, relatively high wear rate, mainly of abrasive-adhesive character was identified. The dependence between wear and implant in situ longevity was not clearly linear, which suggests the influence of other parameters on the polyethylene wear rate. An important role of implant positioning on survival was also revealed. Moreover, it seems that it can be a more important parameter than the characteristics of the patient.

Keywords: total hip arthroplasty, Bicon-Plus cup, retrieval analysis, surface analysis, wear measurement, roughness, deformation, survivorship

Published: June 1, 2017  Show citation

ACS AIP APA ASA Harvard Chicago Chicago Notes IEEE ISO690 MLA NLM Turabian Vancouver
RANUŠA M, GALLO J, HOBZA M, VRBKA M, NEČAS D, HARTL M. Wear and Roughness of Bearing Surface in Retrieved Polyethylene Bicon-Plus Cups. Acta Chir Orthop Traumatol Cech. 2017;84(3):159-167. doi: 10.55095/achot2017/024. PubMed PMID: 28809634.
Download citation

References

  1. Brockett CL, Harper P, Williams S, Issac GH, Dwyer-Joyce RS, Jin Z, Fisher J. The influence of clearance on friction, lubrication and squeaking in large diameter metal-on-metal hip replacements. J Mater Sci Mater M. 2008;19:1575-1579. Go to original source... Go to PubMed...
  2. Cichy Z. [Treatment of dysplastic acetabulum using total hip arthroplasty: our intermediate-term results]. Acta Chir Orthop Traumatol Cech. 2006;73:340-344. Go to original source... Go to PubMed...
  3. Dowson D, Hardaker C, Flett M, Isaac GH. A hip joint simulator study of the performance of metal-on-metal joints - Part II: Design. J Arthroplasty. 2004;19:124-130. Go to original source... Go to PubMed...
  4. Dumbleton JH, Manley MT, Edidin AA. A literature review of the association between wear rate and osteolysis in total hip arthroplasty. J Arthroplasty. 2002;17:649-661. Go to original source... Go to PubMed...
  5. Dungl P, Stedry V, Hajny P. [Medium-term Experience with Zweymuller's Bicon Plus Cup in the Treatment of Postdysplastic Osteoarthritits of the Hip Joint.]. Acta Chir Orthop Traumatol Cech. 2000;67:88-92. Go to PubMed...
  6. Edidin AA, Rimnac CM, Goldgber VM, Kurtz SM. Mechanical behavior, wear surface morphology, and clinical performance of UHMWPE acetabular components after 10 years of implantation. Wear. 2001;250:152-158. Go to original source...
  7. Fulín P, Pokorný D, Šlouf M, Nevoralová M, Vacková T, Dybal J, Kaspřiková N, Landor I. [Analysis of Oxidative Damage to Components Removed from Beznoska/Poldi Total Hip Replacements]. Acta Chir Orthop Traumatol Cech. 2016;83:155-162. Go to original source... Go to PubMed...
  8. Galandáková A, Ulrichová J, Langová K, Hanáková A, Vrbka M, Hartl M, Gallo J. Characteristics of synovial fluid required for optimization of lubrication fluid for biotribological experiments. J Biomed Mater Res Part B Appl Biomater. 2016. Go to original source... Go to PubMed...
  9. Gallo J, Havranek V, Cechova I, Zapletalova J. Wear measurement of retrieved polyethylene ABG 1 cups by universal-type measuring microscope and X-ray methods. Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub. 2006;150:321-326. Go to original source... Go to PubMed...
  10. GOM. GOM mbH, ATOS - Industrial 3D Scanning Technology. Available at: http://www.gom.com/metrology-systems/3d-scanner.html.
  11. Korhonen RK, Koistinen A, Konttinen YT, Santavirta SS, Lappalainen R. The effect of geometry and abduction angle on the stresses in cemented UHMWPE acetabular cups - Finite element simulations and experimental tests. Biomed Eng Online. 2005;4:32. Go to original source... Go to PubMed...
  12. Korovessis P, Repantis T, Zafiropoulos A. High medium-term survivorship and durability of Zweymuller-Plus total hip arthroplasty. Arch Orthop Traum Su. 2011;131:603-611. Go to original source... Go to PubMed...
  13. Kurtz SM, Turner J, Herr M, Edidin AA, Rimnac CM. Assessment of surface roughness and waviness using white light interferometry for short-term implanted, highly crosslinked acetabular components. In: Kurtz SM, Gsell, RA, Martell, J, (eds). Crosslinked and thermally treated ultra-high molecular weight polyethylene for joint replacements. ASTM International, West Conshohocken, 2004, pp.41-56. Go to original source...
  14. Little NJ, Busch CA, Gallagher JA, Rorabeck CH, Bourne RB. Acetabular Polyethylene Wear and Acetabular Inclination and Femoral Offset. Clin Orthop Relat R. 2009;467:2895-2900. Go to original source... Go to PubMed...
  15. McBride A, Flynn J, Miller G, Barnes M, Mackie S. Body mass index and acetabular component position in total hip arthroplasty. ANZ J Surg. 2013;83:171-174. Go to original source... Go to PubMed...
  16. Milosev I, Kovac S, Trebse R, Levasic V, Pisot V. Comparison of ten-year survivorship of hip prostheses with use of conventional polyethylene, metal-on-metal, or ceramic-on-ceramic bearings. J Bone Joint Surg Am. 2012;94:1756-1763. Go to original source... Go to PubMed...
  17. Nečas D, Vrbka M, Urban F, Křupka I, Hartl M. The effect of lubricant constituents on lubrication mechanisms in hip joint replacements. J Mech Behav Biomed. 2016;55:295-307. Go to original source... Go to PubMed...
  18. Nečas D, Vrbka M, Křupka I, Hartl M, Galandáková A. Lubrication within hip replacements - Implication for ceramic-on-hard bearing couples. J Mech Behav Biomed. 2016;61:371-383. Go to original source... Go to PubMed...
  19. Nečas D, Vrbka M, Urban F, Křupka I, Hartl M. The effect of lubricant constituents on lubrication mechanisms in hip joint replacements. J Mech Behav Biomed. 2015;55:295-307. Go to original source... Go to PubMed...
  20. Ottink K, Barnaart L, Westerbeek R, van Kampen K, Bulstra S, van Jonbergen HP. Survival, clinical and radiological outcome of the Zweymuller SL/Bicon-Plus total hip arthroplasty: a 15-year follow-up study. Hip Int 2015;25:204-208. Go to original source... Go to PubMed...
  21. Palousek D, Omasta M, Koutny D, Bednar J, Koutecky T, Dokoupil F. Effect of matte coating on 3D optical measurement accuracy. Opt Mater. 2015;40:1-9. Go to original source...
  22. Parkes M, Myant C, Cann PM, Wong JSS. The effect of buffer solution choice on protein adsorption and lubrication. Tribol Int. 2014;72:108-117. Go to original source...
  23. Patil S, Bergula A, Chen PC, Colwell JR, CW, D'Lima DD. Polyethylene wear and acetabular component orientation. J Bone Joint Surg A. 2003;85:56-63. Go to original source... Go to PubMed...
  24. Pavelka T, Linhart M, Houček P. [Hip joint arthroplasty following surgical treatment of acetabular fracture]. Acta Chir Orthop Traumatol Cech. 2006;73:268-274. Go to original source... Go to PubMed...
  25. Pruitt LA, Chakravartula AM. Mechanics of biomaterials: Fundamental principles for implant design. Cambridge University Press, New York, 2011. Go to original source...
  26. Ranuša M, Gallo J, Vrbka M, Hobza M, Paloušek D, Křupka I, Hartl M. Wear analysis of extracted polyethylene acetabular cups using a 3D optical scanner. Tribol Trans. 2017;60:437-447. Go to original source...
  27. Rossler T, Mandat D, Gallo J, Hrabovsky M, Pochmon M, Havranek V. Optical 3D methods for measurement of prosthetic wear of total hip arthroplasty: principles, verification and results. Opt Express. 2009;17:12723-12730. Go to original source... Go to PubMed...
  28. Schmolders J, Amvrazis G, Pennekamp PH, Strauss AC, Friedrich MJ, Wimmer MD, Rommelspacher Y, Wirtz DC, Wallny T. Thirteen year follow-up of a cementless femoral stem and a threaded acetabular cup in patients younger than fifty years of age. Int Orthop. 2017;41:39-45. Go to original source... Go to PubMed...
  29. Topolovec M, Milosev I. A comparative study of four bearing couples of the same acetabular and femoral component: a mean follow-up of 11.5 years. J Arthroplasty. 2014;29:176-180. Go to original source... Go to PubMed...
  30. Trommer RM, Maru MM, Oliveira Filho WL, Nykanen VPS, Gouvea CP, Archanjo BS, Martins Ferreira EH, Silva RF, Achete CA. Multi-scale evaluation of wear in UHMWPE-metal hip implants tested in a hip joint simulator. Biotribology. 2015;4:1-11. Go to original source...
  31. Veen EJ, Schrier JC, Van'T Riet E, Breslau MJ, Barnaart AF. Outcome of the cementless Zweymuller BICON-PLUS Cup and SL-PLUS stem in the very elderly individuals. Geriatr Orthop Surg Rehabil. 2016;7:74-80. Go to original source... Go to PubMed...
  32. Vrbka M, Návrat T, Křupka I, Hartl M, Šperka P, Gallo J. Study of film formation in bovine serum lubricated contacts under rolling/sliding conditions. P I Mech Eng J-J Eng Tribol. 2013;227:459-475. Go to original source...
  33. Vrbka M, Nečas D, Bartošiík J, Hartl M, Křupka I, Galandáková A, Gallo J. [Determination of a friction coefficient for THA bearing couples]. Acta Chir Orthop Traumatol Cech. 2015;82:341-347. Go to original source... Go to PubMed...
  34. Vrbka M, Nečas D, Hartl M, Křupka I, Urban F, Gallo J. Visualization of lubricating films between artificial head and cup with respect to real geometry. Biotribology. 2015;1-2:61-65. Go to original source...
  35. Wan Z, Boutary M, Dorr LD. The Influence of Acetabular Component Position on Wear in Total Hip Arthroplasty. J Arthroplasty. 2008;23:51-56. Go to original source... Go to PubMed...
  36. Zolotarevova E, Entlicher G, Pavlova E, Slouf M, Pokorny D, Vesely F, Gallo J, Sosna A. Distribution of polyethylene wear particles and bone fragments in periprosthetic tissue around total hip joint replacements. Acta Biomater. 2010;6:3595-3600. Go to original source... Go to PubMed...
  37. Zweymuller KA, Steindl M, Schwarzinger U. Good stability and minimal osteolysis with a biconical threaded cup at 10 years. Clin Orthop Relat Res. 2007;463:128-137. Go to original source...