Interlocked Pins Increase Strength by a Lateral Spread of Load in Femoral Neck Fixation: a Cadaver StudyOriginal papers
- 1 Biomechanics Lab, Division of Orthopaedic Surgery, Oslo University Hospital, Norway
- 2 Institute of Clinical Medicine, Faculty of Medicine, University of Oslo, Norway
PURPOSE OF THE STUDY:
To improve the important torsional, bending and compressive stability in femoral neck fixation, locking plates have been the latest contribution. However, increased strength by restricted fracture motion may come at expense of an altered load distribution and failure patterns. Within locking plate technology, the important intermediate fracture compression may principally be achieved by multiple sliding screws passing through a sideplate fixed to the femur or connected to an interlocking plate not fixed to the femur laterally, sliding "en bloc" with the plate. While biomechanical studies may deliver the short-time patient safety requirements in implant development, no adequate failure evaluation has been performed with interlocking devices ex vivo in this setting. In the present biomechanical study, we analysed if a novel femoral neck interlocking plate with pins could improve fixation performance by changing the parameters involved in the failure mechanism in terms of fixation strength, fracture motion, load distribution and failure pattern.
MATERIAL AND METHODS:
Sixteen pairs of human femurs with stable subcapital osteotomies were fixated by 2 pins or 3 pins interlocked in a plate using a paired design. Femurs were loaded non-destructively to 10° torsion around the neck axis, 200 N anteroposterior bending and 500 N vertical compression in 7° adduction with 1 Hz in 20 000 cycles, and were subsequently subjected to destructive compression to evaluate failure patterns. Bending stiffness, compressive stiffness and displacement from compressive testing reflected fracture motion. Torque and compression to failure replicated known failure mechanisms and defined strength. To evaluate load distribution, associations between biomechanical parameters and measured local bone mineral measurements by quantitative CT were analysed.
RESULTS:
Interlocked pins increased mean strength 73% in torsion and 39% in compression (p = 0.038). Strength was related to all 4 regional mineral masses from the femoral head to subtrochanterically with interlocking (r = 0.64-0.83, p = 0.034), while only to mineral masses in the femoral head in compression and to the head, neck and trochanterically in torsion with individual pins (r = 0.67-0.78, p = 0.024). No difference was detected in fracture motion or failure pattern.
DISCUSSION:
Within the last decade, angular stable implants have expanded our therapeutic arsenal of femoral neck fractures. Increased stability at the expense of altered devastating failure patterns was not retrieved in our study. The broadened understanding of the effect of interlocking pins by an isolated plate as in the current study involved the feature to gain fixation strength. By permitting fracture compression, and through a significant change of correlations between mechanical parameters and local bone mineral factors, a lateral redistribution of load with interlocked pins from the fragile bone medially to the more solid lateral bone was demonstrated. Regarding the long-term patient safety of interlocked pins and healing complications of non-union and segmental collapse of the femoral head, a definite conclusion may be premature. However, the improved biomechanics of an interlocking plate must be considered a favourable development of the pin concept.
CONCLUSIONS:
Interlocked pins may improve fixation performance by a better load distribution, not by restricting fracture motion with corresponding altered failure patterns. This is encouraging and a challenge to complete further studies of the interlocking plate technology in the struggle to find the optimal treatment of the femoral neck fracture.
Keywords: femoral neck fracture, biomechanics, cadaver bone, bone mineral, internal fixation, locking plate, interlocked pins
Published: April 15, 2021 Show citation
ACS | AIP | APA | ASA | Harvard | Chicago | Chicago Notes | IEEE | ISO690 | MLA | NLM | Turabian | Vancouver |
References
- Alshameeri Z, Elbashir M, Parker MJ. The outcome of intracapsular hip fracture fixation using the Targon Femoral Neck (TFN) locking plate system or cannulated cancellous screws: A comparative study involving 2004 patients. Injury. 2017;48:2555-2562.
Go to original source...
Go to PubMed...
- Aminian A, Gao F, Fedoriw WW, Zhang LQ, Kalainov DM, Merk BR. Vertically oriented femoral neck fractures: mechanical analysis of four fixation techniques. J Orthop Trauma. 2007;21:544-548.
Go to original source...
Go to PubMed...
- Bartels S, Gjertsen JE, Frihagen F, Rogmark C, Utvag SE. High failure rate after internal fixation and beneficial outcome after arthroplasty in treatment of displaced femoral neck fractures in patients between 55 and 70 years. Acta Orthop. 2018;89:53-58.
Go to original source...
Go to PubMed...
- Basso T, Klaksvik J, Foss OA. Locking plates and their effects on healing conditions and stress distribution: a femoral neck fracture study in cadavers. Clin Biomech (Bristol, Avon). 2014A;29:595-598.
Go to original source...
Go to PubMed...
- Basso T, Klaksvik J, Foss OA. The effect of interlocking parallel screws in subcapital femoral-neck fracture fixation: a cadaver study. Clin Biomech (Bristol, Avon). 2014B;29:213-217.
Go to original source...
Go to PubMed...
- Bergmann G, Deuretzbacher G, Heller M, Graichen F, Rohlmann A, Strauss J, Duda GN. Hip contact forces and gait patterns from routine activities. J Biomech. 2001;34:859-871.
Go to original source...
Go to PubMed...
- Berkes MB, Little MT, Lazaro LE, Cymerman RM, Helfet DL, Lorich DG. Catastrophic failure after open reduction internal fixation of femoral neck fractures with a novel locking plate implant. J Orthop Trauma. 2012;26:e170-176.
Go to original source...
Go to PubMed...
- Biber R, Brem M, Bail HJ. Targon femoral neck for femoral neck fracture fixation: lessons learnt from a series of one hundred and thirty-five consecutive cases. Int Orthop. 2014;38:595-599.
Go to original source...
Go to PubMed...
- Brandt E, Verdonschot N, van Vugt A, van Kampen A. Biomechanical analysis of the sliding hip screw, cannulated screws and Targon® FN in intracapsular hip fractures in cadaver femora. Injury. 2011;42:183-187.
Go to original source...
Go to PubMed...
- Brattgjerd JE, Loferer M, Niratisairak S, Steen H, Stromsoe K. Increased torsional stability by a novel femoral neck locking plate. The role of plate design and pin configuration in a synthetic bone block model. Clin Biomech. 2018;55:28-35.
Go to original source...
Go to PubMed...
- Eschler A, Brandt S, Gierer P, Mittlemeier T, Gradl G. Angular stable multiple screw fixation (Targon FN) versus standard SHS for the fixation of femoral neck fractures. Injury. 2014;45:S76-S80.
Go to original source...
Go to PubMed...
- FAITH Investigators. Fracture fixation in the operative management of hip fractures (FAITH): an international, multicentre, randomised controlled trial. Lancet. 2017;389:1519-1527.
Go to original source...
Go to PubMed...
- Gjertsen JE, Engesaeter LB, Furnes O, Havelin LI, Steindal K, Vinje T, Fevang JM. The Norwegian Hip Fracture Register: experiences after the first 2 years and 15.576 reported operations. Acta Orthop. 2008;79:583-593.
Go to original source...
Go to PubMed...
- Gjertsen JE, Fevang JM, Matre K, Vinje T, Engesaeter LB. Clinical outcome after undisplaced femoral neck fractures. A prospective comparison of 14,757 undisplaced and displaced fractures reported to the Norwegian Hip Fracture Register. Acta Orthop. 2011;82:268-274.
Go to original source...
Go to PubMed...
- Hendrickson NR, Pickhardt PJ, del Rio AM, Rosas HG, Anderson PA. Bone mineral density T-scores derived from CT attenuation numbers (Hounsfield units): clinical utility and correlation with dual-energy X-ray absorptiometry. Iowa Orthop J. 2018;38:25-31.
Go to PubMed...
- Hunt S, Martin R, Woolridge B. Fatigue testing of a new locking plate for hip fractures. J Med Biol Eng. 2012;32:117-122.
Go to original source...
- Johnson CC, Gausden EB, Weiland AJ, Lane JM, Schreiber JJ. Using Hounsfield units to assess osteoporotic status on wrist computed tomography scans: comparison with dual energy X-ray absorptiometry. J Hand Surg Am. 2016;41:767-774.
Go to original source...
Go to PubMed...
- Kalland K, Åberg H, Berggren A, Ullman M, Snellman G, Jonsson KB, Johansson T. Similar outcome of femoral neck fractures treated with Pinloc or Hansson Pins: 1-year data from a multicenter randomized clinical study on 439 patients. Acta Orthop. 2019;90:542-546.
Go to original source...
Go to PubMed...
- Koval KJ, Sala DA, Kummer FJ, Zuckerman JD. Postoperative weight-bearing after a fracture of the femoral neck or an intertrochanteric fracture. J Bone Joint Surg Am. 1998;80:352-356.
Go to original source...
Go to PubMed...
- Körver RJ, Wieland AW, Kaarsemaker S, Nieuwenhuis JJ, Janzing HM. Clinical experience, primary results and pitfalls in the treatment of intracapsular hip fractures with the Targon FN locking plate, Injury. 2013;44:1926-1929.
Go to original source...
Go to PubMed...
- Meinberg EG, Agel J, Roberts CS, Karam MD, Kellam JF. Fracture and dislocation classification compendium 2018. J Orthop Trauma. 2018;32(Suppl 1):S1-S170.
Go to original source...
Go to PubMed...
- Oakey JW, Stover MD, Summers HD, Sartori M, Havey RM, Patwardhan AG. Does screw configuration affect subtrochanteric fracture after femoral neck fixation? Clin Orthop Relat Res. 2006;443:302-306.
Go to original source...
Go to PubMed...
- Osarumwense D, Tissingh E, Wartenberg K, Aggarwal S, Ismail F. The Targon FN system for the management of intracapsular neck of femur fractures: minimum 2-year experience and outcome in an independent hospital. Clin. Orthop. Surg. 2015;7:22-28.
Go to original source...
Go to PubMed...
- Palm H, Gosvig K, Krasheninnikoff M, Jacobsen S, Gebuhr P. A new measurement for posterior tilt predicts reoperation in undisplaced femoral neck fractures: 113 consecutive patients treated by internal fixation and followed for 1 year. Acta Orthop. 2009;80:303-307.
Go to original source...
Go to PubMed...
- Parker MJ, Gurusamy K. Internal fixation implants for intracapsular hip fractures in adults. Cochrane Database Syst Rev. 2001. http://dx.doi.org/10.1002/14651858 CD001467.
Go to original source...
Go to PubMed...
- Parker MJ, Stedtfeld HW. Internal fixation of intracapsular hip fractures with a dynamic locking plate; initial experience and results for 83 patients treated with a new implant. Injury. 2010;41:348-351.
Go to original source...
Go to PubMed...
- Ragnarsson JI, Kärrholm J. Factors influencing postoperative movement in displaced femoral neck fractures: evaluation by conventional radiography and stereoradiography. J Orthop Trauma. 1992;6:152-158.
Go to original source...
Go to PubMed...
- Ragnarsson JI, Kärrholm J. Stability of femoral neck fracture. Roentgen stereophotogrammetry of 29 hook-pinned fractures. Acta Orthop Scand. 1991;62:201-207.
Go to original source...
Go to PubMed...
- Schemitsch EH, Bhandari M, Boden SD, Bourne RB, Bozic KJ, Jacobs JJ, Zdero R. The evidence-based approach in bringing new orthopaedic devices to market. J Bone Joint Surg Am. 2010;92:1030-1037.
Go to original source...
Go to PubMed...
- Stoffel K, Zderic I, Gras F, Sommer C, Eberli U, Mueller D, Oswald M, Gueorguiev B. Biomechanical evaluation of the femoral neck system in unstable Pauwels III femoral neck fractures: a comparison with the dynamic hip screw and cannulated screws. J Orthop Trauma. 2017;31:131-137.
Go to original source...
Go to PubMed...
- Strömqvist B, Hansson LI, Nilsson LT, Thorngren KG. Hook-pin fixation in femoral neck fractures. A two-year follow-up study of 300 cases. Clin Orthop Rel Res. 1987;218:58-62.
Go to original source...
- Thein R, Herman A, Kedem P, Chechik A, Shazar N. Osteosynthesis of unstable intracapsular femoral neck fracture by dynamic locking plate or screw fixation: early results. J Orthop Trauma. 2014;28:70-76.
Go to original source...
Go to PubMed...
- Wright B, Ragan L, Niratisairak S, Høiseth A, Strømsøe K, Steen H, Brattgjerd JE. High correlation between mechanical properties and bone mineral parameters in embalmed femurs after long-term storage. Clin Biomech. 2018;59:136-142.
Go to original source...
Go to PubMed...
- Yamamoto T, Kobayashi Y, Nonomiya H. Undisplaced femoral neck fractures need a closed reduction before internal fixation. Eur J Orthop Surg Traumatol. 2019;29:73-78.
Go to original source...
Go to PubMed...
- Yin H, Pan Z, Jiang H. Is dynamic locking plate (Targon FN) a better choice for treating of intracapsular hip fracture? A metaanalysis. Int J Surg. 2018;52:30-34.
Go to original source...
Go to PubMed...