Acta Chir Orthop Traumatol Cech. 2023; 90(2):85-91 | DOI: 10.55095/achot2023/012
Type IV Titanium Hypersensitivity: Rare, or Rarely Detected?Current concepts review
- 1 University of Ostrava, Faculty of Medicine, Institute of Emergency Medicine, Ostrava, Czech Republic
- 2 University Hospital Ostrava, Department of Trauma Surgery, Ostrava, Czech Republic
- 3 V©B-Technical University of Ostrava, Centre of Advanced Innovation Technologies, Ostrava, Czech Republic
- 4 V©B-Technical University of Ostrava, Faculty of Mining and Geology, Ostrava, Czech Republic
- 5 University of Ostrava, Faculty of Medicine, Ostrava, Czech Republic
- 6 Institute of Anatomy, University of Silesia in Katowice, Poland
The presented review aims to summarize the current knowledge of hypersensitivity to titanium - a material widely used in medical applications thanks to its exceptional chemical stability, resistance to corrosion, low specific weight and high strength. The hypersensitivity to metals is usually caused by the Type IV immunopathological reaction. Case reports on allergic reactions to titanium are rare but the actual occurrence can be expected to be much higher, especially due to its problematic detection. Although cutaneous patch tests are widely accepted and used for the diagnosis of hypersensitivity of numerous metals (e.g. Ni), it is notoriously unreliable in case of allergies to titanium, which may be associated with the low percutaneous transport of titanium and its salts. The Lymphocyte Transformation Test has superior sensitivity but it remains mostly unknown among clinicians and there are not many laboratories capable of performing it. This review presents numerous case reports indicating, in combination with the above-mentioned facts, that hypersensitivity to titanium should be considered as a possible cause also in non-specific problems associated with titanium implant failure.
Keywords: titanium, allergy, patch test, lymphocyte transformation test
Published: April 1, 2023 Show citation
References
- Alves F, Ribeiro JC, Alves M, Gonçalo M. Titanium allergy as a possible cause of extrusion of a bone conduction ear implant. Contact dermatitis. 2020;83:148-149.
Go to original source...
Go to PubMed...
- Aquino M, Mucci T. Systemic contact dermatitis and allergy to biomedical devices. Curr Allergy Asthma Rep. 2013;13:518-527.
Go to original source...
Go to PubMed...
- Belohlavek J, Belohlavkova S, Hlubocky J, Mrazek V, Linhart A, Podzimek S. Severe allergic dermatitis after closure of foramen ovale with Amplatzer occluder. Ann Thorac Surg. 2013;96:e57-59.
Go to original source...
Go to PubMed...
- Davis MD, Wang MZ, Yiannias JA, Keeling JH, Connolly SM, Richardson DM, Farmer SA. Patch testing with a large series of metal allergens: findings from more than 1,000 patients in one decade at Mayo Clinic. Dermatitis. 2011;22:256-271.
Go to original source...
Go to PubMed...
- de Graaf NPJ, Feilzer AJ, Kleverlaan CJ, Bontkes H, Gibbs S, Rustemeyer T. A retrospective study on titanium sensitivity: Patch test materials and manifestations. Contact dermatitis. 2018;79:85-90.
Go to original source...
Go to PubMed...
- Delgado-Ruiz R, Romanos G. Potential causes of titanium particle and ion release in implant dentistry: a systematic review. Int J Mol Sci. 2018;19:3585.
Go to original source...
Go to PubMed...
- Egusa H, Ko N, Shimazu T, Yatani H. Suspected association of an allergic reaction with titanium dental implants: a clinical report. J Prosthet Dent. 2008;100:344-347.
Go to original source...
Go to PubMed...
- El Khalloufi M, Drevelle O, Soucy G. Titanium: an overview of resources and production methods. Minerals. 2021;11:1425.
Go to original source...
- Engelhart S, Segal RJ. Allergic reaction to vanadium causes a diffuse eczematous eruption and titanium alloy orthopedic implant failure. Cutis. 2017;99:245-249.
Go to PubMed...
- Hettige S, Norris JS. Mortality after local allergic response to titanium cranioplasty. Acta Neurochir. 2012;154:1725-1726.
Go to original source...
Go to PubMed...
- Høl PJ, Kristoffersen EK, Gjerdet NR, Pellowe AS. Novel Nanoparticulate and Ionic Titanium Antigens for Hypersensitivity Testing. Int J Mol Sci. 2018;19:1101.
Go to original source...
Go to PubMed...
- Hosoki M, Nishigawa K, Miyamoto Y, Ohe G, Matsuka Y. Allergic contact dermatitis caused by titanium screws and dental implants. J Prosthodont Res. 2016;60:213-219.
Go to original source...
Go to PubMed...
- Hosoki M, Nishigawa K, Tajima T, Ueda M, Matsuka Y. Cross-sectional observational study exploring clinical risk of titanium allergy caused by dental implants. J Prosthodont Res. 2018;62:426-431.
Go to original source...
Go to PubMed...
- Humble S, Allan Tucker J, Boudreaux C, King JA, Snell K. Titanium particles identified by energy-dispersive X-ray microanalysis within the lungs of a painter at autopsy. Ultrastruct Pathol. 2003;27:127-129.
Go to original source...
Go to PubMed...
- Ishii K, Kodani E, Miyamoto S, Otsuka T, Hosone M, Ogata K, Sato W, Matsumoto S, Tadera T, Ibuki C, Kusama Y, Atarashi H. (2006) Pacemaker contact dermatitis: The effective use of a polytetrafluoroethylene sheet. Pacing Clin Electrophysiol. PACE. 2006;29:1299-1302.
Go to original source...
Go to PubMed...
- Jacobs JJ, Silverton C, Hallab NJ, Skipor AK, Patterson L, Black J, Galante JO. Metal release and excretion from cementless titanium alloy total knee replacements. Clin Orthop Relat Res. 1999;358:173-180.
Go to original source...
- Lalor PA, Revell PA, Gray AB, Wright S, Railton GT, Freeman MA. Sensitivity to titanium. A cause of implant failure? J Bone Joint Surg Br. 1991;73:25-28.
Go to original source...
Go to PubMed...
- Leopold SS, Berger RA, Patterson L, Skipor AK, Urban RM, Jacobs JJ. Serum titanium level for diagnosis of a failed, metal-backed patellar component. J Arthroplasty. 2000;15:938-943.
Go to original source...
Go to PubMed...
- Müller K, Valentine-Thon E. Hypersensitivity to titanium: clinical and laboratory evidence. Neuro Endocrinol Lett. 2006;27(Suppl 1):31-35.
- Nawaz F, Wall BM. (2007) Drug rash with eosinophilia and systemic symptoms (DRESS) syndrome: suspected association with titanium bioprosthesis. Am J Med Sci. 2007;334:215-218.
Go to original source...
Go to PubMed...
- Neta I, Mota M, Lira H, Neves G, Menezes R. Nanostructured titanium dioxide for use in bone implants: a short review. Cerâmica. 2020;66:440-450.
Go to original source...
- Pantejeva T, Nesko A, Macuks M. Case report: Titanium allergy after tendon refixation with suture anchor in a patient with supraspinatus calcifying tendonitis. MEDtube Science. 2017;5:4.
- Peat F, Coomber R, Rana A, Vince A. Vanadium allergy following total knee arthroplasty. BMJ Case Rep. 2018;2018:bcr2017222092.
Go to original source...
Go to PubMed...
- Preez du L, Bütow K, Swart T. Implant failure due to titanium hypersensitivity/allergy? Report of a case. SADJ. 2007;62:22, 24-25.
- Razzante MC, Ehredt DJ, Clougherty CO, Kriger SJ, Menninger BA, Behan Dionisopoulos S, Bhakta PJ, Bruning NG. Type IV cell-mediated hypersensitivity reaction caused by titanium implant following double calcaneal osteotomy and first metatarsal-cuneiform arthrodesis: a case report and review of the literature. J Foot Ankle Surg. 2019;58:974-979.
Go to original source...
Go to PubMed...
- Sakamoto K, Ando K, Noma D. Metal allergy to titanium bars after the Nuss procedure for pectus excavatum. Ann Orthop Surg. 2014;98:708-710.
Go to original source...
Go to PubMed...
- Schalock PC, Dunnick CA, Nedorost S, Brod B ,Warshaw E, Mowad C. American contact dermatitis society core allergen series. Dermatitis. 2013;24:7-9.
Go to original source...
Go to PubMed...
- Sicilia A, Cuesta S, Coma G, Arregui I, Guisasola C, Ruiz E, Maestro A. Titanium allergy in dental implant patients: a clinical study on 1500 consecutive patients. Clin Oral Implants Res. 2008;19:823-835.
Go to original source...
Go to PubMed...
- Singh JA, Yu S, Chen L, Cleveland JD. Rates of Total Joint Replacement in the United States: Future Projections to 2020-2040 Using the National Inpatient Sample. J Rheumatol. 2019;46:1134-1140.
Go to original source...
Go to PubMed...
- Syburra T, Schurr U, Rahn M, Graves K, Genoni M. Gold-coated pacemaker implantation after allergic reactions to pacemaker compounds. Europace. 2010;12:749-750.
Go to original source...
Go to PubMed...
- Thomas P, Bandl WD, Maier S, Summer B, Przybilla B. Hypersensitivity to titanium osteosynthesis with impaired fracture healing, eczema, and T-cell hyperresponsiveness in vitro: case report and review of the literature. Contact Dermatitis. 2006;55:199-202.
Go to original source...
Go to PubMed...
- Thomas P, Thomas M, Summer B, Dietrich K, Zauzig M, Steinhauser E, Krenn V, Arnholdt H, Flaig MJ. Impaired wound-healing, local eczema, and chronic inflammation following titanium osteosynthesis in a nickel and cobalt-allergic patient: a case report and review of the literature. J Bone Joint Surg Am. 2011;93:e61.
Go to original source...
Go to PubMed...
- Van Opstal, N,Verheyden, F. Revision of a tibial baseplate using a customized oxinium component in a case of suspected metal allergy. A case report. Acta Orthop Belg. 2011;77:691-695.
Go to PubMed...
- Vermes C, Kuzsner J, Bárdos T, Than P. Prospective analysis of human leukocyte functional tests reveals metal sensitivity in patients with hip implant. J Orthop Surg Res. 2013;8:12.
Go to original source...
Go to PubMed...
- Wang LF,Wu J,Zheng C, Li SL, Huang RR, Zhang JK. Long-term fever after hallux valgus surgery secondary to titanium allergy: a case report and review of the literature. J Foot Ankle Surg. 2016;55:1282-1286.
Go to original source...
Go to PubMed...